For in vitro laboratory research use only. This article explains analytical chemistry terms. It makes no claims about the effects of any material.
A peptide is a chain of amino acids joined by peptide bonds. Parts of that chain absorb ultraviolet (UV) light, and that is what lets a UV detector see a peptide in HPLC. This guide explains which parts of a peptide absorb UV light, why peptide methods often use 214 nm or 280 nm, how a peptide’s extinction coefficient is estimated from its sequence, and what UV absorbance can and can’t tell you.
What UV absorbance measures
A UV detector or spectrophotometer shines light of one wavelength through a sample and measures how much of it is absorbed. The result is reported as absorbance (A), a number with no units. HPLC software often shows it in milli-absorbance units (mAU).
Absorbance follows the Beer–Lambert law:
- A = ε × c × l
- ε (epsilon) is the molar extinction coefficient, in M⁻¹ cm⁻¹. It describes how strongly a molecule absorbs at that wavelength.
- c is the molar concentration (M).
- l is the path length the light travels through the sample, in cm.
The extinction coefficient depends on the wavelength, so a value is only meaningful with the wavelength it belongs to, written for example as ε280 or ε214.
Why peptide bonds absorb near 214 nm
The amide group in each peptide bond absorbs UV light in the far-UV range. Its absorbance is strongest near 190 nm and falls off through about 210 to 230 nm. Because each peptide has a backbone of peptide bonds, nearly all peptides absorb in this range, even ones with no aromatic amino acids.
Below about 200 nm, many common solvents and oxygen absorb strongly too, which makes the baseline noisy. That is why peptide HPLC methods often detect at around 210 to 220 nm, with 214 nm and 220 nm being common choices. Our guides to the peptide bond and reading an HPLC chromatogram explain the bond and the detector trace.
Which amino acids absorb at 280 nm
At 280 nm the backbone absorbs very little. Absorbance there comes almost entirely from a few side chains:
| Group | Approx. ε280 (M⁻¹ cm⁻¹) | Notes |
|---|---|---|
| Tryptophan (Trp, W) | 5,500 | The strongest absorber at 280 nm |
| Tyrosine (Tyr, Y) | 1,490 | Absorbs most near 275 nm |
| Cystine (one disulfide bond, Cys–Cys) | 125 | Counted per disulfide, not per cysteine |
| Phenylalanine (Phe, F) | Very small | Absorbs mainly near 257 nm, weakly |
A cysteine with a free thiol (–SH) adds almost nothing at 280 nm. Only cysteines paired in a disulfide bond are counted. Our guide to disulfide bonds in peptides explains the difference between free and paired cysteines.
How a peptide’s extinction coefficient is estimated
Because the 280 nm absorbance comes from a few residue types, ε280 can be estimated from the sequence by counting them:
- ε280 ≈ (number of Trp × 5,500) + (number of Tyr × 1,490) + (number of disulfide bonds × 125)
This counting method uses widely published values for each residue type. Keep in mind:
- It is an estimate. Neighbouring residues, pH and the solvent can shift the real value a little. A measured value for a specific sequence is more exact.
- It depends on the disulfide state. The same sequence gives a slightly different value with its cysteines free or paired.
- It needs Trp or Tyr. A peptide with no Trp, no Tyr and no disulfide has an ε280 close to zero, so it gives little or no signal at 280 nm. Far-UV wavelengths (for example around 205 nm), where the peptide bonds absorb, are used to detect these sequences instead.
Side chains absorb at 214 nm too
At 214 nm the peptide bonds are not the only absorbers. Several side chains absorb strongly there as well. One published set of approximate molar values at 214 nm:
| Group | Approx. ε214 (M⁻¹ cm⁻¹) |
|---|---|
| Each peptide bond | 920 |
| Tryptophan | 29,000 |
| Tyrosine | 5,400 |
| Phenylalanine | 5,200 |
| Histidine | 5,100 |
| Methionine | 980 |
| Cysteine | 230 |
This means the 214 nm signal depends on the sequence, not only on chain length. Two different peptides at the same molar amount can give different peak areas. An impurity that is missing an aromatic residue, for example a shorter sequence, can give a smaller peak than its molar amount would suggest, and one that keeps the aromatic residues can give a larger one.
This is one reason HPLC-UV purity is a relative measure. It compares peak areas at the stated wavelength and assumes the components respond to UV light in a similar way. Our guides to HPLC purity and peptide impurities explain how purity is calculated and what the other peaks can be.
214 nm vs 280 nm in HPLC
| Around 214 nm | 280 nm | |
|---|---|---|
| What absorbs | Peptide bonds plus aromatic and some other side chains | Trp, Tyr and disulfide bonds |
| Sees peptides with no Trp or Tyr | Yes | Little or no signal |
| Signal size | Larger | Smaller |
| Baseline | More affected by the mobile phase | Usually flatter |
| Selectivity | Low: most components with peptide bonds appear | Higher: only components with these groups appear |
Trifluoroacetic acid (TFA), common in peptide mobile phases, absorbs at low UV wavelengths. As its amount changes during a gradient, the baseline at 214 nm can drift. Our guides to peptide counter-ions and peptide hydrophobicity cover TFA and retention.
Some detectors, called diode-array or photodiode-array (DAD or PDA) detectors, record a full UV spectrum across each peak. Comparing the spectrum at the start, top and end of a peak can hint that two components are overlapping, but it can’t confirm what either one is.
What UV absorbance can’t tell you
- Identity. Many different sequences absorb at the same wavelengths, so a UV peak doesn’t show which peptide it is. Identity is confirmed by mass, usually with LC-MS. Our guides to LC-MS identity and HPLC vs LC-MS testing explain the difference.
- Net peptide content. Counter-ions and residual water in a lyophilized peptide don’t absorb at 280 nm, so UV absorbance alone doesn’t show how much of a sample’s weight is peptide. Net peptide content is usually measured separately, for example by amino acid analysis. See our guides to net peptide content vs purity and amino acid analysis.
- Components that don’t absorb. Anything that doesn’t absorb at the chosen wavelength won’t appear as a peak and won’t be counted.
A worked example: estimating ε280
Here is an example with made-up numbers, not a real product. Take a made-up peptide whose sequence contains one Trp, two Tyr and two Cys joined in one disulfide bond.
| Contribution | Count | Value (M⁻¹ cm⁻¹) |
|---|---|---|
| Tryptophan | 1 × 5,500 | 5,500 |
| Tyrosine | 2 × 1,490 | 2,980 |
| Disulfide bond | 1 × 125 | 125 |
| Estimated ε280 | 8,605 |
- Because absorbance is proportional to ε, under the same conditions this made-up sequence would give about 1.6 times the 280 nm signal of a made-up sequence with one Trp and no Tyr or disulfide (ε280 ≈ 5,500 M⁻¹ cm⁻¹).
- If the two cysteines were free instead of paired, the estimate would be 8,480 M⁻¹ cm⁻¹, a small difference for this sequence.
- A made-up peptide with no Trp, Tyr or disulfide would have an ε280 near zero and would show little or no signal at 280 nm. It would be detected in the far UV instead.
Frequently asked questions
What wavelength do peptides absorb UV light at?
Peptide bonds absorb most strongly near 190 nm, and peptide methods often detect at around 210 to 220 nm. Peptides containing tryptophan or tyrosine also absorb at about 280 nm.
Why is 214 nm used in peptide HPLC?
Nearly all peptides absorb near 214 nm because of their peptide bonds, and at that wavelength solvents interfere less than they do below about 200 nm.
Why can a peptide show almost no absorbance at 280 nm?
Absorbance at 280 nm comes from tryptophan, tyrosine and disulfide bonds. A sequence without them absorbs very little there.
How is a peptide extinction coefficient calculated?
A common estimate is ε280 ≈ (Trp × 5,500) + (Tyr × 1,490) + (disulfide bonds × 125), in M⁻¹ cm⁻¹. It is an estimate; measured values can differ a little.
Does UV absorbance confirm a peptide’s identity?
No. UV absorbance shows that something absorbing at that wavelength is present. Identity is confirmed by mass, usually with LC-MS.
Does a certificate of analysis state the detection wavelength?
The wavelength is normally stated in the method details. Check which tests a certificate lists.
Peptix certificates
Each Peptix peptide lot with a published certificate has a third-party Certificate of Analysis showing the lot number, HPLC-UV purity and LC-MS identity. Where a certificate reports net content, the measured amount is printed on that certificate. A certificate reports results for the sample tested, so match the lot number on your order to its certificate. To get the certificate for your lot, contact us with the lot number.